MEASUREMENT OF LEDS
(LED Measurement and Terminology)
If there were ever any question that LEDs would be such a dominating
lighting technology just take a look around. They are virtually
everywhere. From standard indicators on stereo equipment, laptops
and toys to traffic lights, variable message signs and automotive
lighting, LEDs have enjoyed an explosive growth over the last
several years with no end in sight.
Much of this is due to the ever-increasing levels of brightness
being achieved with new materials and wafer fabrication processes
as well as the advent of blue and white LEDs for RGB (Full Color)
and general illumination applications. As the level of sophistication
for using LEDs rises, so does the need for accurate measurements
of the LEDs optical properties. I have been in the Optoelectronics
industry for nearly 20 years now and by far the most common type
of question I am always asked is related to the measurement of
an LED. (What is a lumen? How do I convert from lumens to candela?
How do I accurately measure brightness? Why is my measurement
not the same as yours?) I hope to address these and many similar
issues in the subsequent article.
The details of optical measurement in this discussion will be
broken down into four separate but interrelated topics. These
are: Photometric quantities, Radiometric quantities, Wavelength
or Chromaticity quantities and finally Angular or Goniometric
quantities. Although an entire book can easily be written on the
units, standards and test methodologies of the above, I will try
to summarize the more common and basic areas of interest.
Photometry is simply the measurement of light in the visible
spectrum. (approximately 380nm – 770nm) This is light seen
by the naked eye of an average human observer. There are many
different types of photometric units such as nits (cd/m2), lux
(lumen/m2), footcandles (lumen/ft2), stilb (cd/cm2) etc. All of
these are based on two basic photometric standards, the LUMEN
and the CANDELA.
The Candela is the unit of luminous intensity, which can be defined
as the amount of luminous flux (total luminous power emitted from
a source and expressed as lumens) per unit solid angle in a given
direction. The Lumen can be defined as the luminous flux emitted
per unit solid angle from a uniform point source whose luminous
intensity is 1 candela. (1 candela = 1 lumen/steradian) It is
also important to understand the definition of steradian, which
is the solid angle (cone) at the center of a sphere of radius
"r" that subtends an area "r2" on the surface
of the sphere. (See figure 1) The surface area of a sphere is
4p r2; therefore, a sphere has 4p steradians.
FIG 1 - Solid Angle "w" of a Spherical Sector

Most standard LEDs supplied today are measured in candelas, however,
due to the increasing demand for LEDs as a replacement to incandescent
in the general illumination market, the Lumen is now often used
as a unit of measurement for light output. A simple method for
converting from Candela to Lumens is shown in figure 2. Although
empirical calculations are possible for converting many different
types of photometric units, the actual measured value may differ
from the calculated value due to variation in the spatial radiation
characteristics of the LED. In most cases, the empirical calculation
is sufficiently accurate.
FIG 2 - Converting Candelas to Lumens
Step 1.) Obtain the solid angle of the LED
w = p (q )2
w = p (25)2 , assuming the LED half angle is 25º
w = p (.43633)2, convert degrees to radians
w = .598
Step 2.) Calculate Lumens
f = Iv * w
f = 2.00 * .598, assuming the LED brightness is 2000mcd
f = 1.196 Lumens
Additional photometric conversions are shown in figure 3. Although
there are myriad other conversions that are possible, for example
candela to nit or lambert to candela, there is no easy direct
multiplication factor that can be used. Information such as the
area of the source and/or detector, or the measurement distance
and angular characteristics may be required.
Illuminance quantities lux (lx)
footcandle (fc)
phot (ph)
lux (lx) 1
0.0929
0.0001
footcandle (fc) 10.76
1
0.00108
phot (ph) 10,000
929
1
Luminance quantities nit (nt)
stilb (sb)
footlambert (fl)
lambert (L)
apostilb (asb)
nit (nt) 1
0.0001
0.2919
0.00031416
3.1416
stilb (sb) 10,000
1
2919
3.1416
31,416
footlambert (fl) 3.426
0.0003426
1
0.0010764
10.764
lambert (L) 3183
0.3183
929
1
10,000
apostilb (asb) 0.3183
0.0000318
0.0929
0.0001
1
The photometric measurement of LEDs can be more of an art than
an exact science. There are various geometry, electrical and assembly
issues that can greatly affect the optical properties of LEDs.
Because no two LEDs are exactly alike there are steps that will
greatly enhance the accuracy of your measurement. These include
but are not limited to:
- Knowing the LEDs light emission optical center vs.
mechanical center. – When placing LEDs into a
typical test fixture, it is usually placed in such a way as
to assume the light is emanating from the device’s mechanical
center. This is frequently not the case. (See figure 4) The
optical center often deviates 5º or greater from the LED
mechanical center. Although this may not present much of a problem
when measuring devices with a wide viewing angle such as 40º
or greater, for narrow angle devices, the differences in readings
can be considerable. (It should be noted that the CIE –
Commission Internationale de l’Eclairage – recommendation
is to use the mechanical axis of the LED rather than the optical
axis as the measurement reference)
- Measuring the output at a specific time interval or
when stabilized. – When the LED is first energized,
the temperature of the junction increases due to the electrical
power consumed. (The junction temperature of the LED is determined
by Tj = Ta + (Vf * If) * Rth (j-a)) It may take several seconds
or several minutes before the light output has reached thermal
equilibrium and a stabilized value. A decrease in output of
5 –20% or more is not uncommon – This is not permanent
degradation and will recover upon de-energizing. It is often
not practical to wait extended periods of time when many LEDs
require testing, therefore, a set time interval such as 5 seconds
is often established even though the output may not be stabilized.
- Insuring the ambient temperature is consistent during
testing. – LEDs typically change in brightness
and color with temperature. As the temperature rises, the output
decreases and the color shifts towards the higher end of the
spectrum. This will be elaborated upon in the Colorimetry discussion.
- Always use a constant current source. –
The forward voltage (Vf) of an LED can fluctuate from device
to device, therefore, if a standard power supply or voltage
source is used, each LED may not receive the same current.
- Use an easily reproducible test setup. –
Elaborate setups may be fine for laboratory type measurements,
however, when many LEDs require testing, each with different
package styles, viewing angles, colors etc., a system that can
be quickly modified while insuring identical alignment of the
mechanical axis and guaranteeing the detector always sees the
same section of the emission cone is required.
- Insure all equipment is properly maintained and calibrated
FIG 4 – Angle Deviation

Radiometry refers to total radiation or the measurement of all
light whether in the visible, infrared, or ultraviolet spectrum.
The basic unit of radiometric optical power (Radiant Power) is
the watt (W). The watt is an absolute unit because it is independent
of wavelength. One watt of infrared light contains as much power
as one watt of visible light. Other radiometric terms that are
commonly measured are radiant intensity (Watts/Steradian), Irradiance
(W/m2) and Radiance (W/m2 sr). The primary method for measuring
total radiant power/luminous flux is by using an integrating sphere.
(See figure 5)
FIG 5 – Integrating Sphere

The integrating sphere measures light emitted from the LED in
all directions. Generally these measurements are independent of
viewing angle and not subject to angular measurement inaccuracies
seen when testing photometrically, however, errors are still possible.
Sphere diameters of approximately 3 and 6 inches are widely used.
If accuracy is critical, the larger diameter types are preferred
due to the favorable ratio of the sphere area to the size of the
LED and ports, however, this also results in a loss of intensity.
A major source of measurement error has been where to position
the LED inside the integrating sphere. The latest specification
adopted by the CIE , Publication 127, states that the entire package
of the LED should be inside the sphere which is called a 2p luminous
flux measurement.
The same precautions used for measuring LEDs photometrically
should also be followed when making radiometric measurements.
As with photometric conversions, there are a myriad of radiometric
conversions that are possible given the appropriate information.
Radiometric values are normally required for applications used
in conjunction with a photo detector such as in fiber optics,
scanning or sensing.
The scientific measurement and quantification of LED color is
called Colorimetry. Its units are typically given as chromaticity
coordinates or in wavelength. Color perception is very complicated
because it not only depends on the various physical properties
of light but also on things such as surrounding objects, the devices
mechanical properties, the viewer’s eye response as well
as their psychological state. The CIE has established standards
for the measurement of visible light as it relates to the "standard
human eye response." This so called standard observer curve
was first established in 1931 (see figure 6a). From this curve,
the tristimulus values for accurately defining a color are obtained.
The X, Y, Z tristimulus system is based on the assumption that
every color is a combination of three primary colors; red, green
and blue. The 1931 CIE Chromaticity Diagram (see figure 6b) is
derived from the tristimulus values by the following:
x = X/(X+Y+Z) or x = Red/(Red + Green + Blue)
y = Y/(X+Y+Z) or y = Green/(Red + Green + Blue)
Since (x + y + z) =1, the third axis, z = 1 – (x + y)
FIG 6a – 1931 CIE Tristimulus Color Matching Curves

FIG 6b – CIE 1931 Chromaticity Diagram

The chromaticity coordinates are normally specified by the x
and y axis only. In general, most specifications provided by LED
manufacturers do not list the chromaticity coordinates, but rather
the peak and dominant wavelength (unless the LED is white). The
dominant wavelength, specified in nanometers, is obtained from
the color coordinates discussed above. It is essentially the color
that is actually perceived by the human eye.
The peak wavelength is the wavelength at the maximum spectral
intensity. The peak value is easy to obtain and is therefore the
most common value specified by LED manufacturers, however, it
has little practical significance for applications that are viewed
with the human eye since two LEDs may have the same peak wavelength
but can be perceived as different colors.
Currently, the most accurate method for measuring color is by
using a Spectroradiometer. This device performs a complete spectral
power distribution of the source being measured from which all
photometric, radiometric and colorimetric parameters can be mathematically
calculated. The wavelength accuracy of the equipment should be
better than .5nm with .1nm preferred. As mentioned previously,
there are several factors that can affect the value obtained.
One of these is temperature. As the ambient temperature rises,
so to does the LED wavelength. This increase will typically be
from .1nm/ºC - .2nm/ºC depending on the type of LED
used.
The final topic for discussion is goniometric or angular characterization.
A Goniometer is a device that measures the spatial distribution
or radiation pattern of an LED (see figure 8). This can be accomplished
by either moving the detector around the LED or by tilting the
LED while the detector remains stationary. In either case, several
output measurements are taken for each angle as a rotation from
0º – 180º is performed. Upon completion, a profile
of the radiated beam in one plane is obtained. It is often assumed
that since most LEDs are round, the radiation pattern is symmetrical.
This even seems to be indicated by the graphical representations
of viewing angle provided by many LED manufacturers. This is often
not the case. As mentioned previously, the geometric and assembly
variations that occur during the manufacturing of LEDs can greatly
affect its optical properties. It may be necessary to perform
an additional scan and record multiple planes of view. In addition,
some special shaped LEDs such as the oval or elliptical type,
essentially have two radiation patterns (30º x 70º for
example), therefore, both a 0º and 90º scan of the device
is necessary. If a goniometer is not readily available, it is
possible to obtain a crude radiation pattern by using a photo
detector and manually rotating the LED or detector, recording
the output levels and plotting the data points, however, this
can be very tedious and time-consuming.
FIG 8 – Radiation Pattern

It should be clear from what has been discussed that the measurement
of light can be highly inaccurate compared to other more specific
electrical characteristics such as voltage, current or resistance.
There are many factors such as color, device geometry, alignment
of the LED into a test fixture, temperature etc. that can induce
measurement error. It is often classified as being more of an
art than a science. Although a measurement accuracy of ±5%
is still considered standard and widely acceptable in the industry,
with careful attention, accuracies of better than ±2.5%
are possible.
|